The BD® Stem Cell Enumeration (SCE) Assay provides simultaneous enumeration of viable dual-positive CD45+/CD34+ hematopoietic stem cell populations in CD34+ absolute counts (cells/µL) as well as the percentage of the total viable leucocyte count that is CD34+ (%CD34). The BD® Stem Cell Enumeration assay incorporates BD Trucount™ tubes to determine the absolute cell count, thereby eliminating variability associated with hematology-derived absolute counts. Single-platform flow cytometric absolute cell counting protocols have been shown to provide increased robustness of CD34 enumeration by limiting potential sources of imprecision.8 The BD® Stem Cell Enumeration assay incorporates BD Trucount™ tubes to determine the absolute cell count, thereby eliminating variability associated with hematology-derived absolute counts.7
Enumeration of the cell populations in this assay is obtained using either an automated or a manual method for gating and analysis.
The single-tube assay is performed by staining the sample with the reagent in individual BD Trucount™ Tubes for absolute counts.8 When a sample is added to the reagent, the fluorochrome-labeled antibodies in the reagent bind specifically to the cell surface. Additionally, the lyophilized pellet in the BD Trucount™ Tube dissolves, releasing a known number of fluorescent beads. The 7-AAD dye is added to assess viability of the cells. Cells that are 7-AAD+ are not viable. Ammonium chloride is added to lyse erythrocytes before the sample is acquired on a flow cytometer. During analysis of the sample, the concentration of viable CD34+ cells and viable CD45+ cells, and the percentage of viable CD34+ cells in the viable CD45+ cell population, are calculated.